Device not level
The device does not measure weight because it is not leveled. Because tilt degrades measurement accuracy, the Cubis does not allow weighing until it is level.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Place the device on a solid, flat table.
- Start the leveling function in the menu and follow the on-screen guidance.
- Calibrate after leveling.
Possible causes
- The installation surface is tilted
- The device was moved
- The feet were not adjusted
Do you need service?
LEVEL.ERR. clears when the device is leveled. No service is required.
Models showing this code: MCE, Cubis II, Cubis MCE
Source: Sartorius official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.