Code on display COMM.ERR. Sartorius Cubis II MCE

Weighing module communication error

The device cannot get a weight value because there is no communication between the control unit and the weighing module. Weighing is not possible until communication is established.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Wait for the control unit to reestablish communication.
  2. If it does not clear, switch the device off and on again.

Possible causes

  • Temporary loss of communication between the control unit and the weighing module
  • Connection cable problem
  • Electronics failure

Do you need service?

If the communication error recurs, the connection between the control unit and the weighing module must be checked by technical service.

Models showing this code: MCE, Cubis II, Cubis MCE

Source: Sartorius official user manual