Weighing system or electronics error
There is a fault in the weighing system or the device electronics, and the converter signal is not in the normal range. Accurate weighing is not possible in this state.
This fault needs a technician.
What should you do?
You can safely do these while waiting for service.
- Check that the weighing pan is installed correctly, switch the device off and on after 30 seconds. If the error persists, contact service.
- Do not open the device.
- Switch the device off and unplug it until service arrives.
Possible causes
- Weighing system (load cell) failure
- Electronics board failure
- Damage after an overload or impact
Do you need service?
ERR 54 means a weighing system or electronics failure, so it can only be fixed by technical service.
Models showing this code: MCE, Cubis II, Cubis MCE
Source: Sartorius official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.