Review the 26 error and warning messages from the Sartorius Cubis II MCE manufacturer manual, with meanings and fix steps. Device type: analytical balance.

Covered models: MCE, Cubis II, Cubis MCE

LEVEL.ERR. Device not level The weight value is not measured because the device is not level. LOW Weighing pan / underload warning LOW indicates a weighing pan / underload warning on the Sartorius Cubis II MCE. ERR 54 Weighing system or electronics error ERR 54 indicates a weighing system or electronics error on the Sartorius Cubis II MCE. CHK.ERR. Internal memory error An error was detected in the internal storage or firmware memory data of the device. COMM.ERR. Weighing module communication error Communication cannot be established between the control unit and the weighing module. ELISE.ERR. Tilt sensor communication error The device cannot communicate with the tilt sensor. MOTOR.ERR. Motorized leveling error The device cannot level itself automatically. PRT.ERR. Print function error The Print key cannot be used because of the current data interface setting. SYS.ERR. System data error An error was detected in the system data or memory of the control unit. ERR 02 Calibration zero point error Calibration was started without the device being zeroed/tared, or with a load on the weighing pan. ERR 10 Tare function unavailable Manual taring cannot be applied because the active application program is using the tare memory. ERR 11 Value cannot be saved to tare memory The weight value cannot be saved to the tare memory because it is negative or zero. ERR 520 Component or firmware mismatch A component replacement or software update may not have been completed correctly. ERR 521 Component or firmware mismatch A component replacement or software update may not have been completed correctly. ERR 522 Component or firmware mismatch A component replacement or software update may not have been completed correctly. ERR 523 Component or firmware mismatch A component replacement or software update may not have been completed correctly. ERR 524 Component or firmware mismatch A component replacement or software update may not have been completed correctly. ERR 525 Component or firmware mismatch A component replacement or software update may not have been completed correctly. ERR 526 Component or firmware mismatch A component replacement or software update may not have been completed correctly. ERR 527 Component or firmware mismatch A component replacement or software update may not have been completed correctly. APP.ERR. Invalid weight for application The device measured a weight value that is invalid for the application. DIS.ERR. Display format error The output value cannot be shown in the selected display format. HIGH Maximum capacity exceeded The device is overloaded. The maximum weighing capacity has been exceeded. OLDER? Older software update detected The update on the USB drive is older than the software on the device. LOCKED Software update locked A software update cannot be installed on a conformity-approved (legal-for-trade) device. INVALID. No valid update found No valid software update was found on the USB drive.