Code on display ERR 02 Sartorius Cubis II MCE

Calibration zero point error

The device cannot be calibrated because a zero point error occurred when calibration was started. Calibration must start with an empty, zeroed/tared weighing pan.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Remove the sample from the weighing pan.
  2. Tare or zero the device.
  3. If needed, check the preload setting.
  4. Restart the calibration.

Possible causes

  • Taring/zeroing was not done before calibration
  • A load is on the weighing pan
  • The preload setting is not suitable

Do you need service?

Calibration completes after the weighing pan is emptied and zeroed. No service is required.

Models showing this code: MCE, Cubis II, Cubis MCE

Source: Sartorius official user manual