Calibration zero point error
The device cannot be calibrated because a zero point error occurred when calibration was started. Calibration must start with an empty, zeroed/tared weighing pan.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Remove the sample from the weighing pan.
- Tare or zero the device.
- If needed, check the preload setting.
- Restart the calibration.
Possible causes
- Taring/zeroing was not done before calibration
- A load is on the weighing pan
- The preload setting is not suitable
Do you need service?
Calibration completes after the weighing pan is emptied and zeroed. No service is required.
Models showing this code: MCE, Cubis II, Cubis MCE
Source: Sartorius official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.