Code on display LOFSET Rice Lake 120 Plus

Initial load lower limit error

The value measured at power-on is below the initial zeroing range (INIZR) of the calibrated zero point. It usually indicates the platform is missing or has been removed.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Turn the device off and check that the platform is installed correctly.
  2. Make sure nothing under the platform is pushing it upward.
  3. Turn the device on again.

Possible causes

  • Platform not installed or removed
  • Dead load changed since calibration
  • Load cell connection problem

Do you need service?

If the error persists with the platform installed correctly, the zero point or the INIZR setting must be checked by service.

Models showing this code: 120 Plus, 120+

Source: Rice Lake official user manual