Code on display AD HI Rice Lake 120 Plus

A/D positive range error

The signal reaching the analog-to-digital converter is outside the positive measurement range. The manual recommends checking the scale for binding or damage.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Remove the load from the platform.
  2. Check that the platform is not jammed anywhere and is not damaged.
  3. Look for visible damage on the cables.
  4. Switch the device off and on again.

Possible causes

  • Overload
  • Platform jammed
  • Load cell or cable damage

Do you need service?

If the error persists with no load and no binding, there is a load cell or cable fault and service is needed.

Models showing this code: 120 Plus, 120+

Source: Rice Lake official user manual