Parameter or calibration checksum error
The integrity check of the adjustment or calibration data in memory failed. The stored calibration cannot be trusted. According to the manual, recalibration is required.
This fault needs a technician.
What should you do?
You can safely do these while waiting for service.
- Turn the device off and on again after 30 seconds. If the error persists, contact service.
- Do not perform commercial weighing while this error is present.
Possible causes
- Corrupted data in memory
- Electrical interference or sudden power loss
- Memory component failure
Do you need service?
Because the calibration data is corrupted, the device must be recalibrated by technical service.
Models showing this code: 120 Plus, 120+
Source: Rice Lake official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.