Display lower limit
The weight value is too small (too negative) to be displayed, so the lower dashes appear. It may indicate that the zero point has shifted or the platform is not seated properly.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Check that the platform is fully seated.
- Zero the scale with the platform empty.
- Switch the device off and on again.
Possible causes
- Platform removed or not fully seated
- Zeroing performed while loaded
- Load cell connection problem
Do you need service?
If the error persists after zeroing on an empty platform, the load cell or connections must be checked by service.
Models showing this code: 120 Plus, 120+
Source: Rice Lake official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.