Mass or temperature adjustment range error
During mass or temperature adjustment, the final mass or temperature value is outside the permitted range. The adjustment is not saved.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Use the correct nominal weight for the mass adjustment.
- Enter the reference thermometer value correctly for the temperature adjustment.
- Wait for the device to cool down and repeat the adjustment.
Possible causes
- Wrong adjustment weight
- Reference value entered incorrectly for the temperature adjustment
- Heater not cooled down enough
Do you need service?
The error clears when the adjustment is repeated with correct values. No service is needed.
Models showing this code: MAC
Source: RADWAG official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.