Code on display Er1 Hi RADWAG MAC

Initial mass out of range

The initial mass measured during startup and adjustment exceeded the permitted range. The moisture analyzer does not complete startup or adjustment in this case.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Empty the sample pan.
  2. Place the pan handler and the pan correctly.
  3. Restart the device.

Possible causes

  • Sample/load on the pan or pan handler at power-on
  • Pan handler installed incorrectly

Do you need service?

If Er1 persists with an empty and correctly installed pan, technical service is needed for the weighing module.

Models showing this code: MAC

Source: RADWAG official user manual