Code on display Er5 Ad RADWAG MAC

A/D converter error

A/D converter error: the electronic circuit that converts the weighing signal to a digital value is faulty. Moisture measurement cannot be performed in this case.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Turn the device off and on again after 30 seconds. If the error persists, contact service.
  • Do not open the device. The heater section may be hot.
  • While waiting for service, turn the device off and unplug it.

Possible causes

  • A/D converter fault
  • Electronics board failure

Do you need service?

The A/D converter is on the electronics board of the device and can only be repaired by technical service.

Models showing this code: MAC

Source: RADWAG official user manual