A/D upper limit exceeded
The upper limit of the converter (A/D) range has been exceeded. The load on the pan is above the measuring range.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Reduce the amount of sample.
- Check that there is nothing on the pan other than the sample.
Possible causes
- Too much sample added
- Pressing down on the pan or placing a heavy object
Do you need service?
If it persists with an empty pan, technical service is needed for the weighing system.
Models showing this code: MAC
Source: RADWAG official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.