Code on display No Calibration OHAUS Ranger 7000 / R71

No calibration data

Without calibration data the scale cannot calculate weight accurately, so the scale must be calibrated.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check that the calibration menu and the LFT lock are off.
  2. Calibrate the scale with the correct calibration weight.
  • On devices in legal (LFT/approved) use, do not break the seal. The seal must be opened only by technical service.

Possible causes

  • Calibration data deleted
  • Calibration not performed

Do you need service?

On sealed devices, or if the calibration cannot be saved, technical service calibration is required.