Code on display Overload OHAUS Ranger 7000 / R71

Overload

The load on the pan is above capacity. Loading beyond capacity can damage the load cell.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Reduce the load on the weighing pan.
  2. Weigh the load in portions so it stays within capacity.
  • Do not drop the load onto the pan.

Possible causes

  • Load exceeding capacity
  • Shock loading

Do you need service?

If it continues after the load is removed, contact service to check the load cell.

Source: OHAUS official user manual · Open manufacturer document