Code on display Power on Overload OHAUS Ranger 7000 / R71

Overload at power-up

The scale detects a load on the pan at power-up and cannot take the initial zero. If it appears with an empty pan, the calibration may have drifted.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Empty the weighing pan.
  2. Check that nothing is touching the weighing pan.
  3. Switch the balance off and on again.

Possible causes

  • There was a load on the weighing pan at switch-on
  • An object is touching the weighing pan
  • Calibration drift

Do you need service?

If it continues with an empty pan, contact technical service to check the calibration or load cell.

Source: OHAUS official user manual · Open manufacturer document