Tare out of range
If the tare is taken in one unit and the unit is then switched, the converted value may exceed the tare limit. In that case the tare is considered invalid.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Clear the tare.
- Select the unit you will use, then take the tare again.
Possible causes
- Unit changed after taring
- Tare large relative to capacity
Do you need service?
This is a setting limit and does not require service.
Source: OHAUS official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.