Code on display Power on Underload OHAUS Ranger 7000 / R71

Underload at power-up

If the pan is not installed or is seated incorrectly, the power-up signal is lower than expected.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check that the pan is in place and seated correctly.
  2. Switch the balance off and on again.

Possible causes

  • The weighing pan is not installed
  • The weighing pan is seated incorrectly
  • Calibration drift

Do you need service?

If it continues with the pan installed correctly, the calibration or load cell must be checked.

Source: OHAUS official user manual · Open manufacturer document