Calibration error
The calibration is not accepted because the value read during calibration is not within the expected range. It is usually caused by the wrong weight or an unstable environment.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Use the correct calibration weight.
- Move the scale to a place free of vibration and drafts.
- Repeat the calibration.
Possible causes
- Wrong calibration weight
- Unstable environment
Do you need service?
If the calibration is rejected even when repeated with the correct weight, a service check is required.
Source: OHAUS official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.


