Code on display Underload OHAUS MB92

Underload, no pan support

The analyzer does not see the weight of the pan support during use. Weighing is not possible until the pan support is installed.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Re-install the pan support.
  2. Check that the pan support is seated correctly.
  • The heater and sample chamber may be very hot. Take care when opening the cover.

Possible causes

  • Pan support removed
  • Pan support moved out of place

Do you need service?

If Underload appears with the pan support in place, the load cell must be checked by a technician.

Models showing this code: MB92

Source: OHAUS official user manual · Open manufacturer document