Code on display Err 10.5 OHAUS MB92

Temperature unchanged after start-up

The temperature does not rise even though the heater is running. The manual lists unstable mains voltage as a likely cause.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Switch off high-power devices around the analyzer.
  2. Restart the moisture analyzer.
  3. Try the test again.
  • The heater and sample chamber may be very hot. Take care when opening the cover.

Possible causes

  • Unstable mains voltage
  • High-power devices running on the same line
  • Heater or sensor fault

Do you need service?

If Err 10.5 remains with nearby devices switched off, the heater and temperature sensor must be checked by a technician.

Models showing this code: MB92