Heating overshoot above 20 °c
The heater went far above the set temperature. The manual states that this error cannot be fixed by the user.
This fault needs a technician.
What should you do?
You can safely do these while waiting for service.
- Stop the test and let the analyzer cool down.
- Switch the device off and on again.
- The heater and sample chamber may be very hot. Take care when opening the cover.
- Do not leave the analyzer unattended while the temperature is out of control.
Possible causes
- Temperature sensor failure
- Heating control circuit fault
Do you need service?
The temperature sensor and heating control circuit must be checked by a technician.
Models showing this code: MB92
Source: OHAUS official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.