Message on display Temperature calibration parameters error OHAUS MB92

Temperature calibration parameters error

The analyzer cannot read the temperature adjustment values needed to control the heater. The manual states that this error cannot be fixed by the user.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Switch the device off and on again.
  2. Note whether the error appears at every start-up.
  • Do not trust moisture results while Temperature calibration parameters error is shown.

Possible causes

  • Corrupted temperature calibration data
  • Electronics board failure

Do you need service?

The temperature calibration parameters must be reloaded or the board repaired by a technician.

Models showing this code: MB92

Source: OHAUS official user manual · Open manufacturer document