Code on display Err 10.3 OHAUS MB92

Temperature sensor out of range, high

The analyzer cannot read the temperature within its valid range and cannot control heating safely. The manual states that this error cannot be fixed by the user.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Switch the analyzer off and let it cool down.
  2. Switch it on again and check whether the error remains.
  • The heater and sample chamber may be very hot. Take care when opening the cover.
  • Do not open the housing. Internal parts carry mains voltage.

Possible causes

  • Temperature sensor failure
  • Sensor connection problem
  • Electronics board failure

Do you need service?

The temperature sensor and its connection must be checked by a technician.

Models showing this code: MB92

Source: OHAUS official user manual · Open manufacturer document