Code on display Err 68 Minebea Intec X3 PR 5410

No mechanical stability

The weight does not settle during calibration or operation. Vibration, friction, or the filter setting may be the cause.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check that the scale/tank is free and not rubbing against anything.
  2. Stop the sources of vibration.
  3. Repeat the operation when the load has settled.

Possible causes

  • Vibration or swaying load
  • Object or pipe connection rubbing against the scale
  • Unsuitable filter setting

Do you need service?

If there is no mechanical problem, the filter and stability settings must be adjusted by service.

Models showing this code: PR 5410, PR5410/00, PR5410/01, PR5410/03, X3

Source: Minebea Intec official user manual