Code on display Error 7 Minebea Intec X3 PR 5410

Negative input signal

The load cell signal or supply voltage polarity is connected in reverse.

This fault needs a technician.

What should you do?

  • Do not open the device housing. Only technical service should work on internal settings and boards.

Possible causes

  • Load cell signal polarity reversed
  • Supply voltage polarity reversed

Do you need service?

Correcting the cable polarity is a service operation performed in the junction box.

Models showing this code: PR 5410, PR5410/00, PR5410/01, PR5410/03, X3

Source: Minebea Intec official user manual