Code on display Error 2 Minebea Intec X3 PR 5410

Overload

There is a load above capacity at the weighing point or the capacity setting is wrong. In silo and tank systems, overfilling can also cause this error.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Reduce the load on the scale/tank.
  2. Have the capacity (Max) setting checked for compatibility with the system.
  • Stop loading an overloaded tank or platform to avoid the risk of mechanical damage.

Possible causes

  • Load exceeding capacity
  • Wrong capacity setting

Do you need service?

If the error continues after the load is reduced, contact service to check the load cell or the setting.

Models showing this code: PR 5410, PR5410/00, PR5410/01, PR5410/03, X3

Source: Minebea Intec official user manual