Code on display 4-Err KERN & Sohn PNJ

High deviation in ınternal adjustment

The balance compares each internal adjustment with the last one. If the difference is large, the result is not considered reliable.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Leave the balance switched on for its warm-up time.
  2. Eliminate sources of vibration and drafts.
  3. Repeat the internal adjustment.

Possible causes

  • Vibration or air drafts
  • Balance not warmed up
  • Large temperature change

Do you need service?

If 4-Err repeats in stable conditions, the internal adjustment mechanism must be checked by a technician.

Models showing this code: PNJ 600-3M, PNJ 3000-2M, PNJ 12000-1M