Code on display C-Err KERN & Sohn PNJ

Clock error

The clock is used for date, time and timed functions. A faulty clock leads to wrong values in these.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Switch the balance off and on again.
  2. Try to set date and time again in the menu.

Possible causes

  • Fault in the clock circuit
  • Clock battery depleted

Do you need service?

If C-Err remains, the clock circuit must be repaired by a technician.

Models showing this code: PNJ 600-3M, PNJ 3000-2M, PNJ 12000-1M

Source: KERN & Sohn official user manual · Open manufacturer document