Code on display 3-Err KERN & Sohn PNJ

Pan loaded during adjustment

Internal adjustment must be done with an empty pan. If a load is detected, the balance does not complete the adjustment.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Empty the weighing pan completely.
  • Restart the adjustment and do not touch the pan until it ends.

Possible causes

  • Sample or container left on the pan
  • Something placed on the pan during adjustment

Do you need service?

3-Err clears when adjustment is repeated with an empty pan.

Models showing this code: PNJ 600-3M, PNJ 3000-2M, PNJ 12000-1M