Code on display u-Err KERN & Sohn PNJ

Insufficient preload

The balance expects to carry the weight of the pan at all times. If the pan is missing or seated wrongly, the reading falls below the lower limit.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Install the weighing pan.
  2. Check that the pan sits level and touches nothing.
  3. Switch the balance off and on again.

Possible causes

  • The weighing pan is not installed
  • The weighing pan is seated incorrectly

Do you need service?

If u-Err remains with the pan correctly fitted, the weighing system must be checked by a technician.

Models showing this code: PNJ 600-3M, PNJ 3000-2M, PNJ 12000-1M

Source: KERN & Sohn official user manual · Open manufacturer document