Code on display un.oVEr Dini Argeo DGT / DGTP / DGTQ

Printing out of range

A weight that exceeds capacity by 9 divisions, or that is 100 divisions below gross zero, is considered invalid and is not printed.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Reduce the load to the capacity limit.
  2. Zero with ZERO on an empty platform and try again.

Possible causes

  • Load above capacity
  • Platform lifted or zero drift

Do you need service?

If it persists on an empty platform, the calibration or load cell must be checked by service.

Models showing this code: DGTP, DGTPK, DGTQ, DGT60, DGT100, DGT20, DGT1

Source: Dini Argeo official user manual