Alibi memory error
When the indicator cannot find the alibi memory card at startup, it automatically turns on unit conversion mode, but this setting is not saved to SET-UP. Legal weighing records are not stored in this case.
This fault needs a technician.
What should you do?
You can safely do these while waiting for service.
- Turn the device off and on again after 30 seconds. If the error persists, contact service.
- Do not open the device to install or remove the card.
Possible causes
- Alibi memory card not installed
- Communication problem between the card and the indicator
Do you need service?
The alibi memory card is related to legal records, so installing or replacing it is a job for technical service.
Models showing this code: DGTP, DGTPK, DGTQ, DGT60, DGT100, DGT20, DGT1
Source: Dini Argeo official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.