Code on display ------ Dini Argeo DGT / DGTP / DGTQ

Overload / underload

The dashes in the upper row indicate that the weight exceeds maximum capacity by 9 divisions. The dashes in the lower row indicate that the reading has fallen below gross zero (-100 divisions on an approved device, -capacity -9 divisions on a non-approved one).

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Remove the excess load from the platform.
  2. Check that the platform is seated properly.
  3. Zero with ZERO on an empty platform.
  • Do not load the platform above its capacity.

Possible causes

  • Load above capacity
  • Platform dislodged
  • Zero point drift

Do you need service?

If the dashes persist on an empty platform, service is required for a load cell or cable fault.

Models showing this code: DGTP, DGTPK, DGTQ, DGT60, DGT100, DGT20, DGT1

Source: Dini Argeo official user manual