Overload / underload
The dashes in the upper row indicate that the weight exceeds maximum capacity by 9 divisions. The dashes in the lower row indicate that the reading has fallen below gross zero (-100 divisions on an approved device, -capacity -9 divisions on a non-approved one).
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Remove the excess load from the platform.
- Check that the platform is seated properly.
- Zero with ZERO on an empty platform.
- Do not load the platform above its capacity.
Possible causes
- Load above capacity
- Platform dislodged
- Zero point drift
Do you need service?
If the dashes persist on an empty platform, service is required for a load cell or cable fault.
Models showing this code: DGTP, DGTPK, DGTQ, DGT60, DGT100, DGT20, DGT1
Source: Dini Argeo official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.