Capacity point off factory setting
It appears when the calibration knob is returned to the For Use position. A typical cause is an object placed on the balance during capacity point adjustment.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Remove the object from the pan.
- Turn the calibration knob slowly and fully.
- Repeat the span adjustment from the start.
- Do not force the calibration knob or stop it halfway.
Possible causes
- Object placed on the pan during adjustment
- Calibration knob stopped halfway or forced
- Damaged mechanism
Do you need service?
If 4-Err repeats with an empty pan, the internal weight mechanism must be checked by a technician.
Models showing this code: AJ-2200CEN, AJ-3200CEN, AJ-4200CEN, AJ-6200CEN, AJ-8200CEN, AJ-12KCEN, AJH-2200CEN, AJH-3200CEN, AJH-4200CEN
Source: Vibra official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.