Code on display CAL E3 Shimadzu UW

Large drift during PCAL

PCAL adjusts the built-in weight against a certified external weight. The manual advises using the correct weight.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check that your weight matches the value asked on the display.
  2. Place the weight in the centre of the pan.
  3. Repeat PCAL.

Possible causes

  • External weight of wrong value
  • Weight not placed properly
  • Unstable environment

Do you need service?

If the error continues with the correct weight, the internal weight mechanism must be checked by a technician.

Source: Shimadzu official user manual