Code on display CAL E0 Shimadzu UW

Weight loading mechanism error

The manual advises checking the transportation screws. If they are still tight, the internal weight cannot move.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Turn the balance off.
  2. Turn the transportation screws on the bottom anti-clockwise until they stop.
  3. Turn the balance on and repeat calibration.
  • Do not lay the balance on its side while turning the screws.

Possible causes

  • Transportation screws not loosened
  • Internal weight mechanism stuck

Do you need service?

If the error continues with the screws loosened, the internal weight mechanism must be checked by a technician.

Source: Shimadzu official user manual