Shimadzu

UW error codes

21 error codes

Review the 21 error and warning messages from the Shimadzu UW manufacturer manual, with meanings and fix steps. Device type: analytical / precision balance.

CAL E0 Weight loading mechanism error CAL E0 reports a problem in the mechanism that loads the internal calibration weight. CAL E1 Unstable load at calibration CAL E1 reports that the load on the pan was unstable during calibration. CAL E2 Large zero drift at calibration CAL E2 reports that the zero point drifted too much during calibration. CAL E3 Large drift during PCAL CAL E3 reports a large drift during calibration of the built-in weight (PCAL). CAL E4 Large drift at span calibration CAL E4 reports a large drift during span calibration. CAL E5 Wrong calibration weight CAL E5 reports that the weight used for calibration is wrong. CHE x Balance failure CHE x reports a failure in the balance, which stops operating. ComErr Communication command error ComErr reports that the received command code is not correct. dSP oL Display digits exceeded dSP oL reports that the integer part of the result in the selected unit exceeds 7 digits. Err 0x Internal failure Err 0x reports an internal failure in the balance. Err 10 PCAL password error Err 10 reports that the password entered for PCAL is wrong. Err 20 Invalid numeric value Err 20 reports an attempt to set an invalid numeric value. LoCKEd Menu locked LoCKEd reports an attempt to select a menu item while the menu is locked. Err 24 Abnormal supply voltage Err 24 reports that the supply voltage to the balance is abnormal. OL Overload OL reports that the weighing capacity has been exceeded. -OL Negative overload -OL reports that the load is too light or the pan is not set correctly. d ouEr Large error at calibration check d ouEr reports that the calibration check with the built-in weight found too large an error. d UndEr Large error at calibration check d UndEr reports that the calibration check with the built-in weight found too large an error. oFF Reset after power failure oFF reports that the balance restarted after a power failure. Abort Operation aborted Abort reports that the operation was cancelled. wAit Built-in weight moving wAit reports that the built-in calibration weight is moving and you should wait.