Code on display CAL E1 Shimadzu UW

Unstable load at calibration

The balance could not get a steady reading for calibration. The manual advises avoiding wind and vibration.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Move the balance away from drafts and vibration.
  2. Use the windbreak.
  3. Check that nothing touches the pan.
  4. Repeat the calibration.

Possible causes

  • Air drafts
  • Vibrating bench
  • An object is touching the weighing pan

Do you need service?

If the error continues in a suitable place, the weighing mechanism must be checked by a technician.

Source: Shimadzu official user manual