Code on display Er701 Shimadzu MOC-120H

Power supply error

An error was detected in the power supply. According to the manual, the most common cause is that the voltage selector switch on the rear panel is not in the position matching the mains voltage in use.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Turn the device off and unplug it.
  2. Check that the voltage selector switch on the rear panel is in the position matching the mains voltage and correct it if necessary.
  3. Plug into a grounded outlet and turn the device on again.
  • Change the voltage selector switch only while the device is off and unplugged.

Possible causes

  • Voltage selector switch on the rear panel in the wrong position
  • Mains voltage problem

Do you need service?

If Er701 continues despite the correct voltage setting, service must examine the power circuit.

Models showing this code: MOC-120H

Source: Shimadzu official user manual · Open manufacturer document