Power supply error
An error was detected in the power supply. According to the manual, the most common cause is that the voltage selector switch on the rear panel is not in the position matching the mains voltage in use.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Turn the device off and unplug it.
- Check that the voltage selector switch on the rear panel is in the position matching the mains voltage and correct it if necessary.
- Plug into a grounded outlet and turn the device on again.
- Change the voltage selector switch only while the device is off and unplugged.
Possible causes
- Voltage selector switch on the rear panel in the wrong position
- Mains voltage problem
Do you need service?
If Er701 continues despite the correct voltage setting, service must examine the power circuit.
Models showing this code: MOC-120H
Source: Shimadzu official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.