Heater cover / heating system error
The heater cover stayed open for 15 seconds during the measurement. With the cover open, the sample cannot be heated properly and the result is invalid.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Close the heater cover completely.
- Check that the weighing pan and support parts are installed in the correct orientation and that the cover closes without obstruction.
- Do not touch the inside of the cover while the heater is hot.
- Restart the measurement and do not open the cover during it.
Possible causes
- Cover opened during measurement
- Cover not fully closing because parts are installed upside down
- Cover detection mechanism fault
Do you need service?
If Er306 also appears with the cover fully closed, service must check the cover detection system.
Models showing this code: MOC-120H
Source: Shimadzu official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.