Code on display Er306 Shimadzu MOC-120H

Heater cover / heating system error

The heater cover stayed open for 15 seconds during the measurement. With the cover open, the sample cannot be heated properly and the result is invalid.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Close the heater cover completely.
  2. Check that the weighing pan and support parts are installed in the correct orientation and that the cover closes without obstruction.
  • Do not touch the inside of the cover while the heater is hot.
  • Restart the measurement and do not open the cover during it.

Possible causes

  • Cover opened during measurement
  • Cover not fully closing because parts are installed upside down
  • Cover detection mechanism fault

Do you need service?

If Er306 also appears with the cover fully closed, service must check the cover detection system.

Models showing this code: MOC-120H