Code on display Er302 Shimadzu MOC-120H

Temperature sensor error

A fault was detected in the sensor that measures heater temperature. Because the temperature cannot be measured correctly, drying cannot be done safely and accurately.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Turn the device off and on again. If the error continues, contact service.
  • Do not open the device cover.
  • Do not touch the heater until it has cooled down.

Possible causes

  • Temperature sensor failure
  • Break in the sensor connection

Do you need service?

A temperature sensor fault requires part replacement and can only be fixed by technical service.

Models showing this code: MOC-120H

Source: Shimadzu official user manual · Open manufacturer document