Weight too low
The weight placed on the weighing pan during external calibration is much lighter than the expected calibration weight. If calibration is done with the wrong weight, all measurements will be incorrect.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Check the calibration weight value requested on the display.
- Place a calibration weight of the correct nominal value and suitable tolerance in the center of the weighing pan.
- Restart the calibration.
Possible causes
- Calibration weight of the wrong nominal value
- The weight is outside its tolerance
- The weight is not seated properly on the weighing pan
Do you need service?
Calibration completes when the correct calibration weight is used. No service is required.
Models showing this code: Practum
Source: Sartorius official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.