Message on display The balance needs to be adjusted! Sartorius Practum

Calibration / adjustment required

The balance needs to perform a calibration/adjustment (isoCAL). When the set time or a temperature change is exceeded, the balance requests an adjustment to maintain accuracy.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Empty the weighing pan.
  2. Start the calibration/adjustment by pressing the isoCAL / CAL button on the display.
  3. If an external weight is requested, use the calibration weight specified on the display.

Possible causes

  • The isoCAL time interval has expired
  • The ambient temperature changed
  • The balance was moved

Do you need service?

The message clears once the user-initiated calibration/adjustment is done. No service is required.

Models showing this code: Practum

Source: Sartorius official user manual