Message on display The balance needs to perform eCheck! Sartorius Practum

Echeck required

The balance needs to perform its internal self-test (eCheck). eCheck is an internal test that regularly checks whether metrological limits are maintained. If it is not performed, the accuracy of the results is not assured.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Empty the weighing pan and make sure there are no air drafts or vibration.
  2. Start eCheck from the status area on the display.
  3. If eCheck reports a deviation, perform a calibration/adjustment (CAL).

Possible causes

  • The eCheck interval has expired
  • Change in ambient conditions (temperature)

Do you need service?

eCheck is a routine test started by the user. However, if the test fails repeatedly, a service check is required.

Models showing this code: Practum

Source: Sartorius official user manual