Weight too high
The weight placed on the weighing pan during external calibration is much heavier than the expected calibration weight. Calibration with the wrong weight distorts all measurements.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Check that there is nothing on the weighing pan other than the calibration weight.
- Use the correct calibration weight of the value requested on the display.
- Restart the calibration.
Possible causes
- Calibration weight of the wrong nominal value
- Another object is on the weighing pan
- A container or its tare was left on the weighing pan
Do you need service?
Calibration completes when the correct calibration weight is used. No service is required.
Models showing this code: Practum
Source: Sartorius official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.