Code on display LH RADWAG WTC 600

Initial mass error

Initial mass error at startup: the indicator is outside the permitted range of the stored initial mass. According to the manual, the most common cause is a loaded weighing pan at startup.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Empty the weighing pan. The zero indication appears once it is empty.
  2. Check that the weighing pan is installed correctly and restart the balance.

Possible causes

  • Load on the weighing pan at startup
  • The weighing pan is not installed
  • Load cell damage

Do you need service?

If LH continues with the weighing pan empty and installed correctly, technical service is needed for the initial mass or load cell.

Models showing this code: WTC 600, WTC

Source: RADWAG official user manual · Open manufacturer document