Initial mass error
Initial mass error at startup: the indicator is outside the permitted range of the stored initial mass. According to the manual, the most common cause is a loaded weighing pan at startup.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Empty the weighing pan. The zero indication appears once it is empty.
- Check that the weighing pan is installed correctly and restart the balance.
Possible causes
- Load on the weighing pan at startup
- The weighing pan is not installed
- Load cell damage
Do you need service?
If LH continues with the weighing pan empty and installed correctly, technical service is needed for the initial mass or load cell.
Models showing this code: WTC 600, WTC
Source: RADWAG official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.