Code on display null RADWAG WTC 600

Zero value from converter

The A/D converter is returning a zero value, so the weighing system is not producing a signal. It can appear when the weighing pan is not installed, otherwise it points to a load cell or electronics fault.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check that the weighing pan is installed correctly.
  2. Turn the balance off and on again after 30 seconds.
  • Do not open the device.

Possible causes

  • The weighing pan is not installed
  • Load cell connection problem
  • A/D converter fault

Do you need service?

If it continues after restarting with the weighing pan installed correctly, technical service is needed for the load cell or electronics.

Models showing this code: WTC 600, WTC

Source: RADWAG official user manual · Open manufacturer document