Code on display Err4 RADWAG WTC 600

Calibration weight / starting mass out of range

The calibration (adjustment) weight is outside the ±1% tolerance, or the starting mass is outside the ±10% tolerance. The balance does not accept the adjustment in this case, and an adjustment with the wrong weight distorts measurements.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Empty the weighing pan and check that it is installed correctly.
  2. Use the adjustment weight of the nominal value the balance requests.
  3. Restart the adjustment.

Possible causes

  • Adjustment weight of the wrong nominal value
  • The weighing pan was not empty during adjustment
  • The weighing pan is not installed or the starting mass changed

Do you need service?

If Err4 persists even with the correct weight and an empty weighing pan, technical service is required for the load cell/calibration.

Models showing this code: WTC 600, WTC

Source: RADWAG official user manual · Open manufacturer document