Code on display Err8 RADWAG WTC 600

Process timeout

The tare, zeroing, initial mass determination or adjustment process could not be completed in time. The balance waits for a stable reading for these operations, and vibration or air drafts prevent stability.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Move the balance to a sturdy table free of vibration.
  2. Eliminate air drafts (air conditioner, window).
  3. Check that the weighing pan is not touching anything and repeat the process.

Possible causes

  • Vibration or air drafts
  • Wobbly table
  • Weighing pan touching something

Do you need service?

If Err8 continues even in a stable environment, the weighing system needs technical service.

Models showing this code: WTC 600, WTC

Source: RADWAG official user manual · Open manufacturer document